5.4 Replacement parts
Description
Maintenance kit, central cleaning brush
Section 6 Chloride sensor (007254)
6.1 Specifications
Specifications are subject to change without notice.
Specification
Range
Accuracy
Resolution
Parameters
Deployment life of sensor cartridge
4
Life
of sensor cartridge
Maximum depth
Operating temperature
6.2 Product overview
The chloride sensor (007254) is green. Refer to
There is a replaceable sensor cartridge installed in the sensor. Refer to
the deployment life of the sensor cartridge.
The sensor cartridge has a depth limit of 15 meters (49 feet). Remove the sensor cartridge and install
the sensor cap before the sonde is deployed deeper than 15 m (49 ft) to prevent damage to the
sensor cartridge. Refer to the illustrated steps in
To extend the life of the sensor cartridge, remove the sensor cartridge and install the sensor cap
when the sensor is not in use. Refer to the illustrated steps in
Figure 6 Chloride sensor
1 Sensor
2 Sensor cartridge
4
Deployment life plus the storage life
Details
0 to 18,000 mg/L
± 10% or ± 5 mg/L (the larger value)
0.01 mg/L
–
Cl
or sensor mV
1 to 2 years
1 to 2 years
15 m (49 ft)
0 to 50 °C (32 to 122 °F)
Figure
6.
Figure
7.
Figure
7.
3 Sensor cap
4 Glass vial for sensor cartridge storage
Item no.
9480900
Specifications
on page 15 for
English 15