Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 5

General Features

Power Supply 100÷240 Vac
Display 4 digits with 7 segments
Bnc connection for probe pH
Switch 3 position Priming, ON, Stand by
Peristaltic pump flowrate and pressure look at label on product
Measure Range pH: 3.0÷8.0 pH (resolution: 0.1 PH)
Dosage method ON-OFF ( Set Point adjustable)
Calibration method pH 1 Point
Switch position
:
1) Priming position "the pump running" (Switch position MOM)
2) Normal position "the instrument measure pH and dosing. (Switch position ON)
Keep Press
key for 5 seconds to run calibration menu:
pH probe: the pump displays C7pH press
Displays quality probe ex: C100
Keep Press
key until flashing Setpoint value press again to increase the value
when you are ready to leave a
The Setpoint values can be set from 4.0 until 7.0 pH increase step by step 0.1 pH,
When the real measure is less than 3.0 pH Flashing the value 3.0
When the real measure is more than 8.0 pH Flashing the value 8.0
3) Stand by position "The pump is stop" (Switch position OFF)
Keep Press
key for 3 seconds to run Setup menu:
pH setup menu:
1] Adjust dosage method for pH Acid or Alka
2] Temperature fixed Value 25°C
Kontrol 01
Kontrol 01
key and wait 60 seconds
key after 3 seconds the unit save the value and flashing mem.
code: 0000137199 rev. 4.0

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

68264

Tabla de contenido