Technical Data - Air Liquide Soffio Cube Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 18
all deposits are washed away and leave to dry.
Disinfecting
The compressor unit (1) can be disinfected with a clean cloth and some chemical disinfectant
following dosage and use limitations as provided by disinfectant manufacturer.
All the accessories (except the connecting tube) can be disinfected with chemical disinfectants
following dosage and use limitations as provided by disinfectant manufacturer. Disinfectants
are usually available at pharmacies.
Steam sterilizing
The nebuliser and its accessories (except air hose and masks) can be heat steam sterilized
up to 121 °C (20 min.) or 134 °C (7 min.). EN554/ISO11134.The sterilization packaging must
conform to EN868/ISO11607 and be suitable for steam sterilization. After sterilization always
let all components cool down to ambient temperature before further use.
Do not repeat sterilization cycle when components are still warm.
4 - MAINTENANCE
The SOFFIO CUBE aerosol therapy device does not require any particular maintenance.
The compressor does not need to be lubricated as the piston is the dry-operating type. It is
advisable replace nebulizer after some 100 to 120 treatments on single patient or after about
20 sterilization cycles. Check the filter continually for cleanliness and replace it if dirty, or after
a maximum of 3 months use. Spare filters are provided with the device. To replace the filter,
open the air filter slot (4) (if needed using a tool) and replace the filter (Fig. D).
5 - PROBLEMS, POSSIBLE CAUSES AND SOLUTIONS
The device
cannot be
switched on
The nebuliser
functions poorly
or not at all
If you have failed to solve the problem after performing the above checks and the device still
does not work, have it checked by an authorized service centre.

6 - TECHNICAL DATA

Supply voltage
Operation
GLISH • ENGLISH • ENGLISH • ENGLISH • ENGLISH • ENGLIS
Ensure the power cord (2) is correctly plugged into the socket.
Ensure the ON/OFF switch (3) is in the position «I».
Ensure that the connecting tube (14) is correctly connected at
both ends. Ensure the connecting tube is not squashed, bent,
dirty or blocked. If necessary, replace with a new one. Ensure
the nebuliser is correctly assembled. Ensure the required
medication has been added and in the appropriate quantity.
Ensure that the small hole of the nebulizer's lower body (9) is
not clogged.
230 V ~ 50 Hz
continuous

Publicidad

Tabla de contenido
loading

Tabla de contenido