PROBLEM
Printer not
functioning
Results printed in
duplicate
Lot number and
assay specificity not
on printout
Bubbles in the buffer
tubing
No fluid being
delivered by the
pipette
IF A PROBLEM CANNOT BE SOLVED, PLEASE REFER TO SUPPLIER.
Insert Code: EIN023
POSSIBLE
?
CAUSE(S)
1. Printer not connected
2. Printer settings
incorrect
When printer is
connected to
MININEPH
, results
PLUS
print automatically
New reagent lot details
not loaded
Buffer not primed
Tubing detached from
connectors
Tubing split
Page 22 of 121
SUGGESTED
ACTION
Ensure that the printer is connected
to the MININEPH
PLUS
plug (RS232)
Ensure printer settings are correct
(see section 4.4.1)
Do not press PRINT button unless a
duplicate printout is required
Press ESC button, follow
instructions, press 2=N for lot
number check, swipe card and
repeat assay
Re-start the assay session and
select „1 = Prime‟ when the option is
offered. Always refill the buffer bottle
or use a fresh bottle of buffer before
priming.
If bubbles are persistent and are not
removed by 2 primes then agitate
the tubing while priming. To remove
bubbles from line A, agitate line A
while the syringe is moving in a
downward direction. To remove
bubbles from line B or from the
pipette line agitate the tubing while
the syringe is moving in an upward
direction.
Check all tubing connections and
replace any detached lengths of
tubing (see section 7)
Replace the length of tubing (see
section 7)