Cleaning The Battery - Stryker SMRT 6500 Manual De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 217

CLEANING THE BATTERY

English
WARNING
• In addition to personal protective equipment (PPE), always wear insulative
rubber gloves when cleaning the SMRT Pak to reduce the risk of injury.
• While power washing, use any appropriate personal safety equipment
(goggles, respirator, etc.) to avoid the risk of inhaling contagion. Use of power
washing equipment can aerate contamination collected during the use of the
cot.
The SMRT Pak is designed to be power washable. The preferred method for
power washing the SMRT Pak is with the standard hospital surgical cart washer
or hand held wand unit.
CAUTION
• DO NOT STEAM CLEAN OR ULTRASONICALLY CLEAN THE SMRT PAK.
• Maximum water temperature should not exceed 180°F/82°C.
• Maximum air dry temperature (cart washers) is 240°F/115°C.
• Maximum water pressure should not exceed 1500 psi/130.5 bar. If a hand
held wand is being used to wash the unit, the pressure nozzle must be kept
a minimum of 24 inches (61 cm) from the SMRT Pak.
Return To Table of Contents
1-26
Cleaning
6500-109-105 REV A
www.stryker.com

Publicidad

Tabla de contenido
loading

Tabla de contenido