Drive Status Menu
This menu can be accessed without any access password. It
displays various technical parameters regarding the real-time
working conditions of Turbo-Smart. Listed below is the more
commonly required information.
Blower Output Frequency
(max. setting for version A: 85 Hz
max. setting for version B: 110 Hz)
Blower Output Voltage
(max. setting: 220V)
Blower Overall Current
(max. setting for version A: 5,5 A
max. setting for version B: 7,5 A)
Pump (Centrifuge) Output Frequency
(fixed setting for versions A and B: 55 Hz)
Pump (Centrifuge) Output Voltage
(max. setting: 220V)
Pump (Centrifuge) Overall Bus Current
(max. setting for version A: 5,5 A
max. setting for version B: 7 A)
System Temperature
Maximum Detected Temperature
(+58 °C: Alarm and Stop of the aspirator)
Vacuum
(max. setting: 200 mbar)