Autoclavable "U" Contra Angle - 20,000 rpm
Directions for Use
0297
75-12
Precautions
• This is a stationary head type instrument; users
should not remove the head from the sheath.
• Do not put angle through ultrasonic disinfection
solution cleaning procedure.
• Do not use cold sterilization or dry heat.
• The contra angle must be lubricated before
every autoclaving procedure.
• Damage due to lack of lubrication is misuse
and is not covered by the warranty.
• Maximum autoclave temperature is 270°F.
Lubrication Instructions
1. A er thoroughly cleaning the outside of the angle
with a disinfectant moistened gauze sponge,
open the latch and insert two drops of oil into hole
"A" and hole "B".
2. Close the latch and rotate the sha several times
to circulate the oil. Wipe off any excess oil expelled
from the oil holes.
3. Sterilize by autoclaving before pu ing back
into service.
Cleaning/Sterilization Instructions
1. Thoroughly clean the outside of the angle with a
disinfectant moistened gauze sponge.
2. Autoclave angle per manufacturer's instructions.
The contra angles have been tested with no
adverse effects at 250° F (121° C) with a 30
minute exposure time.
Symbols
Caution: See warning or precautions
Consult instructions for use and website address
Manufacturer
Authorized Representative in the
EC REP
European Union
Catalog number
LOT
Lot number
Complies with requirements of
EU Directive 93/42/EEC
0297
Caution: Federal (USA) law restricts this device to
sale by or on the order of a dentist or physician.
Integra York PA, Inc
589 Davies Drive, York, PA 17402 USA
.
866-854-8300 USA 717-840-2763 outside USA 717-840-9347 fax
integralife.com/integra-miltex
Integra LifeSciences Services (France) SAS Immeuble Sequoia 2
EC REP
97 allée Alexandre Borodine
Parc Technologique de la Porte des Alpes
69800 Saint Priest–France
Telephone: 33 (0) 4 37 47 59 10
Fax: 33 (0) 4 37 47 59 29
Integra and the Integra logo are trademarks of, and Miltex is a registered
trademark of, Integra LifeSciences Corporation or its subsidiaries.
© 2011 Integra LifeSciences Corporation. All rights reserved.
DENT7512DFU Rev. C 06/13
Integra
Miltex
®
Grado medio
Contra ángulo en "U" apto para autoclave - 20.000 rpm
Instrucciones de uso
0297
75-12
Precauciones
• Este es un tipo de instrumento de cabezal fijo; los
usuarios no deben quitar el cabezal de la funda.
• No coloque el ángulo en el procedimiento de
limpieza ultrasónica con solución desinfectante.
• No use esterilización en frío o calor seco.
• El contra ángulo debe lubricarse antes de cada
procedimiento de esterilización en autoclave.
• Los daños debidos a la falta de lubricación, la cual
está considerada como una mala utilización, no
están cubiertos por la garantía.
• La temperatura máxima de esterilización en
autoclave es de 132 ºC (270 °F).
Instrucciones de lubricación
1. Después de limpiar completamente la parte exterior
del ángulo con una esponja de gasa mojada en
desinfectante, abra el enganche e introduzca dos
gotas de aceite en el orificio "A" y en el orificio "B".
2. Cierre el enganche y gire el eje varias veces para
hacer circular el aceite. Limpie cualquier exceso de
aceite que salga de los orificios para el aceite.
3. Esterilice con autoclave antes de volver a ponerlo
en servicio.
Instrucciones de limpieza/esterilización
1. Limpie completamente la parte exterior del ángulo con
una esponja de gasa mojada en desinfectante.
2. Esterilizar el ángulo en autoclave siguiendo las
instrucciones del fabricante. Los contra ángulos se han
probado con un tiempo de exposición de 30 minutos a
121 °C (250 °F) y no se han encontrado efectos adversos.
Símbolos
Precaución: Véase advertencia o precauciones
Consulte las instrucciones de uso y la
dirección del sitio web
Fabricante
Representante autorizado en la Unión Europea
EC REP
Número de catálogo
LOT
Número de lote
Cumple los requisitos de la directiva de
la UE 93/42/EEC
0297
Precaución: Las Leyes Federales de EE. UU.
restringen la venta de este dispositivo por
prescripción de un dentista o médico.
Integra York PA, Inc
589 Davies Drive, York, PA 17402 USA
.
866-854-8300 USA 717-840-2763 outside USA 717-840-9347 fax
integralife.com/integra-miltex
Integra LifeSciences Services (France) SAS Immeuble Sequoia 2
EC REP
97 allée Alexandre Borodine
Parc Technologique de la Porte des Alpes
69800 Saint Priest–France
Telephone: 33 (0) 4 37 47 59 10
Fax: 33 (0) 4 37 47 59 29
Integra and the Integra logo are trademarks of, and Miltex is a registered
trademark of, Integra LifeSciences Corporation or its subsidiaries.
© 2011 Integra LifeSciences Corporation. All rights reserved.
DENT7512DFU Rev. C 06/13
Integra
Miltex
®
Mid Grade
Contre-angle autoclavable en « U » - 20 000 rpm
Mode d'emploi
0297
75-12
Précautions
• C'est un instrument de type à tête fixe ; les
utilisateurs ne doivent pas retirer la tête de la gaine.
• Ne pas pencher lors de la procédure de nettoyage
par solution de désinfection aux ultrasons.
• Ne pas utiliser de stérilisation à froid ni de
chaleur sèche.
• Le contre-angle doit être lubrifié avant toute
procédure d'autoclave.
• Tout dommage dû au manque de lubrification est
une mauvaise utilisation non couverte par la
garantie.
• La température maximale d'autoclave est de
132 °C (270 °F).
Instructions de lubrification
1. Après avoir ne oyé soigneusement l'extérieur de
l'angle avec une compresse de gaze imbibée de
désinfectant, ouvrir le verrouillage et insérer deux
gou es d'huile dans les trous « A » et « B ».
2. Fermer le verrouillage et faire pivoter la tige
plusieurs fois pour faire circuler l'huile. Essuyer tout
excès d'huile expulsé des trous d'huile.
3. Stériliser par autoclave avant de reme re en service.
Instructions de ne oyage/stérilisation
1. Ne oyer soigneusement l'extérieur de l'angle avec
une compresse de gaze imbibée de désinfectant.
2. Stériliser l'angle par autoclave en respectant les
instructions du fabricant. Les contre-angles n'ont
pas subi d'effets secondaires lors des tests à 121°C
(250°F) avec un temps d'exposition de 30 minutes.
Symboles
A ention : Consulter les avertissements ou
les précautions
Consulter le mode d'emploi et visiter
le site Internet
Fabricant
Représentant agréé en Europe
EC REP
Référence
LOT
Numéro de lot
Conforme aux exigences de la
directive 93/42/CEE
0297
A ention : Aux États-Unis, la loi fédérale
réserve la vente de ce dispositif aux médecins
ou sur prescription médicale.
Integra York PA, Inc
589 Davies Drive, York, PA 17402 USA
.
866-854-8300 USA 717-840-2763 outside USA 717-840-9347 fax
integralife.com/integra-miltex
Integra LifeSciences Services (France) SAS Immeuble Sequoia 2
EC REP
97 allée Alexandre Borodine
Parc Technologique de la Porte des Alpes
69800 Saint Priest–France
Telephone: 33 (0) 4 37 47 59 10
Fax: 33 (0) 4 37 47 59 29
Integra and the Integra logo are trademarks of, and Miltex is a registered
trademark of, Integra LifeSciences Corporation or its subsidiaries.
© 2011 Integra LifeSciences Corporation. All rights reserved.
DENT7512DFU Rev. C 06/13

Publicidad

loading

Resumen de contenidos para Integra Miltex 75-12

  • Página 1 Fax: 33 (0) 4 37 47 59 29 Fax: 33 (0) 4 37 47 59 29 Integra and the Integra logo are trademarks of, and Miltex is a registered Integra and the Integra logo are trademarks of, and Miltex is a registered Integra and the Integra logo are trademarks of, and Miltex is a registered trademark of, Integra LifeSciences Corporation or its subsidiaries.
  • Página 2 Fax: 33 (0) 4 37 47 59 29 Fax: 33 (0) 4 37 47 59 29 Integra and the Integra logo are trademarks of, and Miltex is a registered Integra and the Integra logo are trademarks of, and Miltex is a registered Integra and the Integra logo are trademarks of, and Miltex is a registered trademark of, Integra LifeSciences Corporation or its subsidiaries.