OPERATION
• Wear protective eyewear and ensure glare shield is secured in place before use.
• Light guide can become hot during use, avoid contact with it.
On / Off
1. Press the "Start/Stop" button (A) once to turn on the display panel. The digital display will
show the curing time from the previous cycle.
2. Press the "Start/Stop" button to begin or stop the curing cycle.
3. Two (2) beeps will countdown the last 2 seconds of the cycle and a 3rd beep will
sound at the end of the cycle.
Note: The device will automatically switch off after being left at rest for 30 seconds. Press the
"Start/Stop" button to turn the power back on.
Setting curing time
1. To program curing mode, press and hold the "Mode" button (B) until the display flashes.
2. Push the "Mode" button to increase the time in 5-second increments. When the time reaches
15 seconds, pressing the "Mode" button again results in the display rolling over to the
minimum setting of 5 seconds.
Press
Time Setting
1
5 Seconds
2
10 Seconds
3
15 Seconds
3. Press the "Start/Stop" button (A) to enter the new curing mode.
Screen displays
HH (Overheating): The handpiece is equipped with a sensor that will
automatically turn off the light to prevent overheating. This may occur during
long periods of high use. The handpiece will remain off for 6 minutes to allow
for complete cooling.
Note: The EnLITE is not suitable for whitening applications.
bL (Low Battery): When the battery power is low, "bL" will appear in the
display and the handpiece will start beeping. Place the handpiece in the
charger to recharge battery.
A
Mode
Full Power
Ramp Up
Pulse
B
3