If the overfill security system fails, continued aspiration will be stopped by the bacteriological filter.
Should this back-up security also fail, there is a risk of penetration of liquid into the device. Don't attempt to proceed with aspiration under these
circumstances.Return the device for service.
The manufacturer will provide upon request electric diagrams, components list, descriptions, setting instructions and any other
information that can help the technical assistance staff for product repair.
The Manufacturer cannot be held liable for accidental or indirect damages should the device be modified, repaired without authorization
or should any of its component be damaged due to accident or misuse.
Any minimal modification / repair on the device voids the warranty and does not guarantee the compliance with the technical
requirements provided by the MDD 93/42/EEC (and subsequent changes) and its normatives.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, PLEASE CONTACT TECHNICAL SERVICE.
THE MANUFACTURER DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL SERVICE CHECKING, APPEARS TO BE
•
Place the unit on a flat, horizontal surface
•
Connect the end of the short silicon tube (1), with antibacterial filter (8), to the suction connector (2).
•
The other tube already connected to the filter has to be connected to the "VACUUM" jar outlet (3), where has been fixed the red float (security
float). When the 90% of the volume of the jar is reached there is the activation of the security float (the float close the aspiration connector on
the jar) to avoid liquid penetration inside the device.
WARNING: Ensure that the FLUID SIDE or IN marker on the filter is on the side facing the collection jar lid and fitted into the "VACUUM".
A wrong connection causes immediate destruction in case of contact with sucked liquids.
•
Connect the long silicone tube (4) to the "PATIENT" jar outlet (5)
•
Connect the other end of the long silicon tube (4) to the probe plastic connector (6) then connect the suction probe to it.
•
Connect the power cord to the device then connect the plug to the electrical mains supply.
•
Push switch (7) on position I to start suction.
•
Unscrew the jar's lid and fill the jar 1/3 full or ordinary water (this for an easy cleaning operations and an rapid reaching of the functionality
vacuum) then rescrew the lid on the jar correctly.
•
During operation the jar has to be in vertical position to avoid overflow valve to cut off aspiration. Should this happen, switch off the device and
disconnect the tube from the jar cover (from "VACUUM" outlet).
•
Once finished push switch on O position and unplug.
•
Remove the accessories and clean.