Mettler Toledo WKC Manual De Usuario página 8

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 20
4 Commissioning
1 Switch on the weigh module at the final location (place of use) only.
2 Observe the warm-up time (at least 30 minutes after power up) before starting with the weighing operation.
3 When putting the weigh module into operation for the first time, perform an internal adjustment to adjust the
weigh module for the place of use.
4 Open the APW-Link software and connect the weigh module. (Instead of the APW-Link software, the MT-
SICS Command C3 can be used.)
5 Select Test & Adjustment from the Home Tab.
6 Start the Adjustment process with the internal or an
external weight and follow the instructions on the
screen.
6
Commissioning​ ​
WKC

Publicidad

Tabla de contenido
loading

Tabla de contenido