4. Installation Instructions

To open the wash chamber door, plug in the unit,
turn it on and press the unlock symbol.
Tools and supplies required for installation:
• Slot screwdriver
• Adjustable wrench
Ensure that HIP Cleaning Solution (instrument wash
chemical) is available. All other supplies are included
with the Hydrim unit.
Specifications:
Voltage:
220 – 240 V
Frequency: 50 Hz
Rated load: 2 kW
Circuit
breaker:
16 A
Running Noise: 60 dBA
Utility Hook-ups
Hose /
Length /
Cord
Diameter
Hot inlet
1.9 m
G 3/4"
Cold Inlet
1.9 m
G 3/4"
Drain
1.5 m
3/4"
Electrical
1.8 m
AWG 18-3
power cord
The electrical outlet should not be located directly behind the unit and should be accessible.
• Pliers
Height, built-in: 830mm
Height, free-standing: 850 mm
Width:
600 mm
Depth:
600 mm
Depth with
door open:
1200 mm
Weight:
80 kg
Max. Distance
Water
Distance from
Pressure
inlet / drain
1.5 m
2-5 bar
1.5 m
2-5 bar
cold water inlet hose

Installation Configurations:

If the Hydrim is installed in a Sterilisation Center,
the manufacturer of the Sterilisation Center should
allow a 10mm space at the top, back and both
sides of the Hydrim. This will facilitate installation,
levelling, and service access to the Hydrim.
Do not move the Hydrim into place by maneuver-
ing the open wash chamber door. This may cause
to door to become misaligned and leak.
Levelling the Hydrim:
The unit is standing on three supports: rollers
(wheels) at the back and two legs at the front.
1. Remove the kickplate. Push the Hydrim into
place while lifting the strap at the front to allow
the unit to roll on the rollers.
2. Adjust the front legs as required until the Hydrim
is level. Access the legs from inside the unit.
3. The rear two legs are used only if the floor is
uneven or cannot provide support to the rollers.
Shut
-off
valve
Yes
Yes
hot water inlet hose
35
drain hose
rollers