7. Troubleshooting
Error Messages
Error
CF1 Water Heating
failure
CF2 Chamber
Filling Failure
CF3 Chamber
temperature
reading failure
CF4 Water
evacuation
failure
CF5 Disinfection
Failure
CF6 Serial
Communication
Failure
CF8 Secondary
Chamber Temperature
Sensor Failure
CF9 Software or
PCB Failure
CF10 Pressure Sensor
Failure
CF13 Temperature
Validation Error
CF14 Flow Meter
Error
CF15 Chamber
overflow
CF16 No pressure
Cause
The water does not reach the
required temperature in the
specified time
The water does not fill the
chamber within the specified time.
Faulty sensor.
Water cannot drain
from the unit.
Temperature dropped below
disinfection temperature during
the disinfection phase.
Failure to read the second
temperature sensor within 10 sec.
Second temperature sensor
readings are out of limits.
Cycle length is more than 3hr.
Faulty sensor.
Second temperature sensor
readings are out of limits.
Detergent dosing validation error
Overflow switch did not turn off after
30 sec of running the drain pump.
Water pressure drops during
washing or disinfection.
56
Solution
Call for service.
Check that the water hoses
are not kinked. Open the
water shut-off valves.
Call for service.
Check that the drain hose is
not kinked. Check that the
filters in the bottom of the
chamber are not obstructed.
Call for service.
Call for service.
Call for service.
Call for service.
Call for service.
Call for service.
Call for service.
Check if drain is blocked;
Call for service.
Call for service.