Error Codes - IKA Heizung C 1.20 Instrucciones De Manejo

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 27
Turn the C 1 calorimeter off and on again; at this point, the hea-
ter will be recognized by the C 1 calorimeter.
The error message "Heater not
NOTICE
found" should not be displayed
(see "Error codes" section).
 Open the water tap slowly.
Check connections for tightness.
CAUTION
 Check the cooling water temperature on the C 1 calorimeter in
the "Accessories" menu under the "Condenser test" item.
 Then choose the corresponding measuring mode for the C 1
calorimeter in the "settings" menu (see "Use for the intended pur-
poses" section).
If the water inlet temperature is
NOTICE
too high, select a higher tempe-
rature range for the C 1 to allow
a measurement to be performed.
If the water inlet temperature is
NOTICE
low, you should always select the
lower operating temperature for
the C 1 to save time and energy.
Proceed as follows in the case of error code:
 Switch the device off,
 Perform corrective measures as described below,
 Perform a restart.
Error
Cause
Heater not
C 1.20 heater is not recognized by
found
the C 1 calorimeter.
 Perform a system test in accordance with the C 1 calorimeter
operating instructions or a measurement.

Error Codes

Effect
No heater
If the actions described fail to resolve the fault or another error
code is displayed then take one of the following steps:
- Contact the service department,
- Send the device for repair, including a short description of the
fault..
Fig. 20
Solution
Always turn the heater on first.
Check the heating connection cable plug contact.
Switch the C 1 calorimeter off and on again.
Fig. 18
Fig. 19
16