English - HEINE EN 50 UNPLUGGED Instrucciones De Uso

Ocultar thumbs Ver también para EN 50 UNPLUGGED:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 15
HEINE
EN50 UNPLUGGED
®
HEINE
mPack UNPLUGGED
®
Please read and follow these instructions for use of and keep
them for future reference.
Intended Use
The HEINE
mPack UNPLUGGED battery is intended exclusively
®
for supplying power to HEINE instruments on headbands specially
designed for the purpose. There are two chargers available for
charging the Lithium-polymer battery inside the HEINE
UNPLUGGED:
a) the HEINE
EN50 UNPLUGGED, with two charging cradles, is
®
intended for wall mounting.
b) the HEINE
UNPLUGGED plug-in transformer is intended as a
®
charger for mobile operation.
It should only be operated by qualified medical personnel within
a professional healthcare facility.
For U.S. only:
Federal law restricts this device to sale by or on the order of a
Physician or Practitioner.
Warnings and Safety Information
CAUTION! Indicates potential hazardous situations. Ignoring
the corresponding instructions may lead to dangerous situations
of mild to moderate extent. (Background color yellow;
foreground color black).
NOTE! This symbol indicates valuable advice in terms of set up,
operation or maintenance, as applicable. Notes are important,
but not related to hazardous situations.
Product overview
HEINE
mPack UNPLUGGED
®
Connector for the instrument
or an UNPLUGGED power unit
for mobile charging
Capacity indicator
Serial-No.
Charging contacts for charging
via the HEINE
EN50 UNPLUGGED
®
HEINE
EN 50 UNPLUGGED
®
Charging cradle II
Contact pins
Charging cradle I
HEINE UNPLUGGED power unit (plug-in transformer)
Setting up
Mains adapters
EU
USA / Japan
mPack
®
reverse view
Mains adapters
UK
Australia
med 2014
The mains adapter is interchangeable. This allows a world wide use.
The procedure of changing the plugs is as per the pictures.
Take care to mount the charging station outside the patient
environment.
There must be adhered to the distances (1.5m) in accordance
with IEC60601-1 (see figure).
Fig. 1: Minimum dimensions of the patient environment
The device should never be operated or even stored at places
listed below, because this could lead to operating failures:
a. which are heavily exposed to moisture or where water
condensing may occur
b. which are exposed to special environmental conditions
c. which are subject to constant vibrations
d. which are subject to high temperature fluctuations
e. outdoors
Wall mounting the HEINE
Use the four provided screws to fix the mounting bracket to
the wall. Place the EN50 UNPLUGGED on top of the mounting
racket and fasten with the two top screws.
2019-06-05
EN50 UNPLUGGED
®

ENGLISH

7/48

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Mpack unplugged

Tabla de contenido