ulsonix Pro-clean 0.7 Manual De Instrucciones página 9

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 22
GENERAL
USAGE
AND
CLEANING
INSTRUCTIONS
Cleaning tips
Simple cleaning
Just put a few drops of liquid soap in water. You can use any
liquid cleaner that is not acid based.
Complete cleaning
Use proprietary cleaners to remove rust or other intense
stains.
Caution
The use of liquid acid-based cleaners or alkali can lead to
the formation of rust on
the basin walls and even lead to the formation of holes in
the walls of the unit. Be sure to use products at low pH.
PREPARATION
1.
Before using the device ensure that the power sup-
ply complies with the technical requirements on the
device label.
2.
Place the device on a flat, even surface. Fill the tank
with water ensuring that the objects to be cleaned
are completely submersed. Place the objects into
the basket provided.
Warning
Fill the tank up to the 2/3 mark with water. Never oper-
ate the device when empty or with insufficient water! This
might damage the device.
BASIC / BASIC ECO
1.
Heating ON/OFF; heating setting
2.
Indicator for heating in process
16
3.
Timer ON/OFF; Timer setting
4.
Indicator for ultrasonic cleaning in process
OPERATION
1.
Once the switch is rotated clockwise, the machine
will enter into the time operating mode. The time
can be freely modified in accordance with the
indication at the edge of rotary switch on the panel.
2.
The green work indication lamp will turn on and
the ultrasonic will send out the working sound of
"squeak" and the machine will start working.
3.
The cleaning machine will turn off automatically
when time is up.
4.
Rotate the switch anticlockwise to start long-term
operating mode. The infinity mark ∞ is indicated in
the panel.
5.
Green working indicator lamp is on and the ultra-
sonic sends out the working sound of "squeak" and
starts working and will continue working.
6.
When cleaning is completed, in order to power off
please rotate the switch clockwise to the "off" man-
ually.
7.
Rotate the heating switch clockwise to start heating
the cleaning fluid. The heating temperature can be
selected in accordance with the indication at the
edge of rotary switch on the panel.
8.
The green heating indicator lamp will turn on and
the cleaning fluid will be heated.
9.
When the temperature reaches the set value, the
green heating indicator lamp will turn off and heating
will stop.
10.
When the temperature of the fluid will be lower
than set value on the rotary switch on the panel,
then device will start heating again. This way, the de-
vice will maintain the set temperature.
11.
To stop heating, the rotary switch must be turned to
OFF position anticlockwise.
Rev. IV 2017
min. 2/3
MEMORY QUICK / MEMORY QUICK ECO
u l t r a s o n i c c l e a n e r
6
7
9
5
8
4
10
2
3
1
min. 2/3
1.
On/Off button
2.
START/PAUSE & Memory button
3.
Temperature control buttons
4.
Heating process LED
5.
Current temperature indicator for the cleaning
liquid
6.
Set temperature indicator for the cleaning liquid
7.
Time indicator: displays the set time or remaining
time in minutes
8.
Cleaning process LED
9.
Remaining cleaning time counter
10.
Cleaning time adjustment buttons
USE
1.
All buttons are tactile. A simple touch is enough.
Press the button No. 1 to turn on. The cleaning
process begins. You can now set the desired
temperature and time.
2.
Temperature set: Display No 5 shows the
temperature of the cleaning liquid. To heat
the liquid, use arrows No. 3 (up to raise the
temperature, down to lower it). If the temperature
is higher than the current temperature, the unit
starts to heat up and the LED lights up. Once the
desired temperature is reached, the diode turns
off. To terminate the heating process, adjust the
current temperature setting to '00'.
3.
Time set: When the device is on, you see on display
number 7 the time set in the previous cleaning.
Press No 10 to set the time (up to increase, down
to lower)
4.
After setting the time, press START (No. 2) to
begin the cleaning. A red LED turns on. When the
countdown begins (No. 9), the cleanup is almost
Rev. IV 2017
complete. The cleaning will stop automatically at the
end of programmed time. The time programmed by
the last use appears (No. 7).
5.
Pause function and memory: Press the green button
Start / Pause (No. 2) to activate the memory
function. When the time is up and you touch the
green button, the programmed time is saved. If the
cleaning process is underway and you press the
green button, the cleaning is paused. To restart,
repress the green button.
min. 2/3
6.
When cleaning is finished, press the ON / OFF
(No. 1) button to turn off the device. Remember to
unplug it after use. Empty the basin so that the unit
u l t r a s o n i c c l e a n e r
proclean 0.7D
is completely empty.
LCD DEGAS
4
5
3
2
6
1
1.
On/Off button
2.
START/PAUSE & Memory button
3.
Degas button
4.
LCD display
5.
Up/Down control buttons
6.
Set button
OPERATION
1.
All the buttons are touch buttons which only re-
quire a light touch. To activate the device and start
the cleaning process, touch the power button (#1)
at the front of the device. You can then select the
required temperature and time settings.
2.
Setting the temperature: The current temperature
is shown on the LCD display. The display will flash
when you press the Set button (#6) and you can
now set the required temperature by pressing the
17