Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions
and voltage variations on the
power supply input lines
IEC 61000-4-11
Power frequency (50 Hz)
magnetic field
IEC 61000-4-8
Note:
U
is the AC mains voltage prior to application of the test level
T
Guidance and manufacturer's declaration - electromagnetic immunity
The Kanmed BabyWarmer BW3 system is intended for use in the electromagnetic environment specified below. The customer or the user of the
Kanmed BabyWarmer BW3 system should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2,5 GHz
Note 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
Note 2:
structures, objects and people.
* Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To access the electromagnetic environment due to fixed
RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Kanmed
BabyWarmer BW3 system is used exceeds the applicable RF compliance level above, the Kanmed BabyWarmer BW3 system should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the
Operatherm OP3.
** Over the frequency range of 150 kHz to 80 MHz, field strength should be less than 3 V/m.
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV differential mode
±2 kV common mode
<5% U
T
(>95% dip in U
)
T
for 0,5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 sec
3 A/m
Compliance
level
3 Vrms
3 V/m
Página 27 of 28
Manual de Usuario BW3-077/3
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV differential mode
±2 kV common mode
<5% U
T
(>95% dip in U
)
T
for 0,5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 sec
3 A/m
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be
used no closer to any part of the Kanmed BabyWarmer BW3
system, including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
d = 1,17 √ P
d = 1,17 √ P
80 MHz to 800 MHz
d = 2,33 √ P
800 MHz to 2,5 GHz
Where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer and d is
the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey*, should be less than the
compliance level in each frequency range**
Interference may occur in the vicinity of equipment marked with
the following symbol:
at least 30%
Mains power quality should
be that of a typical
commercial or hospital
environment
Mains power quality should
be that of a typical
commercial or hospital
environment.
If continued operation during
power mains interruptions, it
is recommended that the
Kanmed BabyWarmer BW3
system be powered from an
uninterruptible power supply
unit (UPS)
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment