17. Test Results

Summary of Test Results
Test
Documentation (IEC 60601-1-2 sections 4 and 5)
General requirements
General requirements
for EMC.
External labels
Conformity of Users'
Manual
Accuracy of Technical
Description
Emission (IEC 60601-1-2 section 7)
Conducted emission
Freq. range:150 kHz -
30 MHz
Radiated emission
Freq. range: 30 - 1000
MHz
Harmonic current
emission test
Voltage changes,
Voltage fluctuations and
Flicker test
Immunity (IEC 60601-1-2 section 8)
Immunity from
Electrostatic discharge
(ESD)
Immunity from radiated
electromagnetic fields
Immunity from Proximity
field from wireless
communications
equipment
Immunity from Electrical
Fast transient (EFT)
Immunity from Surge
Immunity from
conducted disturbances
induced by radio-
frequency fields
Immunity from voltage
dips, short interruptions
and voltage variations
Standard
Class/ Severity Level
section 4.1
--
section 5.1
--
section 5.2.1
--
section 5.2.2
--
CISPR 11
Group 1 Class B
230, 120 & 100 VAC mains
CISPR 11
Group 1 Class B
IEC 61000-3-2
230 VAC mains
IEC 61000-3-3
230 VAC mains
IEC 61000-4-2
8 kV contact discharges &
15 kV air discharges
IEC 61000-4-3
10.0 V/m; 80 MHz ÷ 2.7 GHz,
80% AM, 1 kHz
IEC 61000-4-3
List of frequencies
from 9 V/m up to 28 V/m,
PM (18 Hz or 217 Hz), FM 1 kHz
IEC 61000-4-4
± 2.0 kV;
Tr/Th – 5/50 ns, 100 kHz
on AC mains
IEC 61000-4-5
±1.0 kV DM;
Tr/Th – 1.2/50 (8/20) µs
on AC mains
IEC 61000-4-6
3.0 V
& 6.0 V
RMS
mains;
0.15÷ 80 MHz, 80% AM 1 kHz
IEC 61000-4-11
0 % - 0.5 cycle & 1 cycle;
70% - 25 cycles; 0% - 250 cycles
on 230 & 100 AC mains
21
21
Test
Result
Complies
Complies
Complies
Complies
Complies
Complies
Test NA
Complies
Complies
Complies
Complies
Complies
Complies
on AC
Complies
RMS
Complies