KPS -TL320 Manual De Instrucciones página 21

Medidor de resistencia de tierra digital
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

● Maximum Operating Error
Operating error (B) is an error obtained within the
rated operating conditions, and calculated with the
intrinsic error (A), which is an error of the instrument
used, and the error (En) due to variations.
B=±(│A│+1.15
A: Intrinsic error
E
: Variation due to changing the supply voltage
2
E
: Variation due to changing the temperature
3
E
: Variation due to series interference voltage
4
E
: Variation due to resistance of the probes and
5
auxiliary earth electrode resistance
● Range to keep the maximum operating error
Measurement range within which the maximum
operating error (±30%) applies.
20Ω Range
200Ω Range
2000Ω Range :
● Temperature & Humidity
Operating : 0°C~50°C ≤ 80%R.H.
Storage : -10°C~60°C ≤ 80%R.H.
2
E
+E
2
:
5 ~ 19.99Ω
:
20 ~ 199.9Ω
200 ~ 1999Ω
-6-
2
2
2
+E
+E
)
3
4
5