PERFORMANCE AND TECHNICAL SPECIFICATIONS
Ambient Operating Temperature
Range
Operating Altitude
Patient Temperature Range
Temperature Predict Time*
Transport/Storage Temperature
Calibration Accuracy
Humidity
Dimensions
Weight
Power
Equipment Classification
IPXØ
* Predict time depends on probe placement and patient condition. Predict
times do not include tissue contact-detection time or time needed to place
the probe into the measurement site.
50.0 °F to 104.0 °F (10.0 °C to 40.0°C)
-557 ft. to +16,000 ft / -170m to +4877m
80.0 °F to 110.0 °F (26.7 °C to 43.3 °C)
Oral:
Approximately 4–6 secs.
Adult axillary (18 years and older):
Approximately 12–15 secs.
Pediatric axillary (17 years and younger):
Approximately 10–13 secs.
Rectal:
Approximately 10–13 secs.
-13°F to +131°F (-25°C to +55°C)
+/- 0.2 °F (0.1 °C) (Monitor Mode)
15% to 95% non-condensing
8.46" x 3.18" x 2.43"
(215mm x 81mm x 62mm)
12.6 ounces/357 grams
(3) 1.5Vdc AA batteries
Class 2 Device, Continuous Operation
Not AP or APG equipment
Not protected against the ingress of water.
1-23