Sunoptic Surgical SSL-3736-2 Manual De Usuario página 6

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 33
The charge process will start after inserting a battery pack into the slot. The charge indication will
be displayed (see Table 1).
LED Indicator
One time Red/Orange/Green
Red/Green blinking
Orange blinking
Green light
Red blinking
Red light
NOTE:
If a red LED is lit on a bay, the battery of the corresponding bay should be removed from
the charger and its contacts checked. Install the battery pack again. If the red LED
illuminates again, remove the battery pack from the charger and contact Sunoptic
Technologies
The unit will become warm during normal operation. This does not indicate a malfunction.
Prevent metallic objects from coming into contact with the metal parts of the DC plug of
the power supply. If this happens, a short may occur and the unit may be damaged.
Ensure the battery has proper orientation before insertion. Do not attempt to force a
battery into the slot as this may damage the contacts.
LIT-232 Sunoptic Surgical
Rev. B
Self-test; charger is ready for use
The battery was not recognized as a Smart Battery. Either a
conventional battery is inserted or an extremely discharged
Smart Battery. If it is a Smart Battery, it will be reactivated
within 15 minutes and recharged. If this is not, the case, the
LED will light red – see below.
The battery is currently being calibrated
The battery is charged and ready for use
The battery is too hot or too cold to be charged without
damage. If the battery is too cold, it will be charged as soon
as it has warmed up sufficiently. If the battery is too hot, it
should be removed to cool down.
Either:
-
The battery is badly damaged and must be replaced
-
It is a conventional battery which cannot be
recharged
Table 1. LED Indicator Legend
®
for service.
®
Page 6 of 48
Charging Condition
(English)

Publicidad

Tabla de contenido
loading

Tabla de contenido