travel with
The
system is designed with universal voltage. That means that you can
use the system all over the world. You will need to use an adaptor for the plug,
since electric wall outlets vary by country. Make sure that
get wet while traveling. Do not use it in a wet environment. It is recommended
to protect your
separately at www.mepower.com.
troubleshooting
Indicator
No lights are on
All
energy setting
buttons are blinking
(High-Medium-Low)
activation button
on the applicator is
blinking
Power button is
blinking in orange
color
No flashes are coming
out
23
EN
device while traveling. A travel bag may be purchased
What does it mean
Device
Power cable is not
connected
No
energy level was
selected
energy level was
selected but
activation
button was not pressed yet
An error had been
detected
1. RF silver skin contact bars
not in contact with the skin
in treatment area
2.
activation button is
not pressed or
energy
level was not selected
system does not
How to react
Connect power cable
Select an
level. The applicator fan
will start blowing indicating that the
cartridge is active
Press the
activation button.
The applicator fan will start blowing
indicating that the
Reset the system by turning it off and on
again. If problem persists, disconnect
power cord and wait until all lights are off
(several seconds). Then reconnect the
power cord and turn on system. If error
light is still on, try moving to a cooler area
with a room temperature of less than
30ºC/ 86ºF and restart the system
In order for the system to flash pulses,
make sure all 3 conditions are met:
level selected,
activation button is
pressed, RF silver skin contact bars are in
contact with skin
cartridge is active