If the user wishes to program any of these specific parameters, he must first enter into USER SETUP by
keeping F2 pressed while the instrument is on.
KEYS FUNCTIONS
In this setting the keys take on the following meanings:
ON/OFF KEY: exits from a step or setup menu storing the modifications.
ZERO/TARE: enters in a step and confirms a value.
F1:- scrolls backward through the steps
- increases the selected value
F2: scrolls forward through the steps
- scrolling of the digits to be entered from left to right.
Step
Display shows
ModE:
FUNCTIONING MODE
tot
norM
PEAK
HoLd
MoFF :
AUTO TURN-OFF
M 00
M 01
M 02
M ...
M 30
irMd:
REMOTE CONTROL CONFIGURATION
ir 4
ir 1
CALF:
FINE CALIBRATION
This step allows to correct a possible maximum calibration error; do the following:
Unload the scale and press the ZERO/TARE key; the display shows the message "donE" for an instant
-
Load the calibration mass and press the F1 and F2 keys to adjust the displayed weight unitl one obtains
-
an indication corresponding to the value of the load placed on the instrument.
Carry out the adjustments observing the à0ß, ~, and NET indicators. The instrument is well calibrated
-
when the ~ indicator is blinking.
Careful: the F1 and F2 keys work on the small corrections. It may be necessary to press the keys various
times so that one can see the corrections on the displayed value of the load.
Press the ON/OFF key
-
USER SETUP
= display functioning + accumulator function
= display functioning
= display functioning + peak detector function
= display functioning + HOLD function
= auto switch-off disabled
= auto switch-off after 1 minute
= auto switch-off after 2 minutes
= auto switch-off after 30 minutes
4 KEYS (see table page 7)
1KEY (ZERO/TARE)
8