Clinician Remote Control Directions for Use

CP Mode: Stimulator

The Remote Control can be used to establish communication between the implanted Stimulator (IPG)
and Clinician Programmer (CP) for device programming. The Stimulator can be placed into CP Mode
(Pairing Mode) by following the steps below. Alternatively, from the Lock Screen, press and hold the P
button (approximately 5 seconds) until the screen displays "Stimulator Ready for CP Connection". This
will place the Stimulator into CP Mode (Pairing Mode) so that the CP may connect to it.
Clinician Menu:
From the
CP Mode: Stimulator.
1. Select
2. Connect the CP to the Stimulator and proceed with device programming.
to exit CP Mode.
3. Press
Note: CP Mode (Pairing Mode) must be enabled for the CP to connect to and program the
Stimulator using Bluetooth communication.

Exit Clinician Mode

From the Clinician Menu:
Exit Clinician Mode. The System Settings screen displays.
1. Select
Clinician Remote Control Directions for Use
92395488-02
28 of 1077