Before initial use the product should be kept in its original packaging. GCE
recommends use of the original packaging (including internal sealing bag
and caps) if the product is withdraw from operation (for transport, storage).
Statutory laws, rules and regulations for medical gases, accident preven-
tion and environmental protection must be observed.
EN
OPERATING CONDITIONS
In case of storage at a temperature below -20 °C, d o not operate the
regulator until it has been allowed to increase its temperature to a mini-
mum of -20 °C.
At regulators intended for use with mixture of medical gases O
lower operation temperature limit +5°C. During normal use the fl ow outlet
and pressure outlet will sometimes have a frosty appearance. This is a
normal physical reaction in the valve, due to that the gas is going from
high pressure to low pressure (Joule Thompson e ect). Ensure that all
equipment that is connected to the valve by at least a 2 metre hose.
4. PERSONNEL INSTRUCTIONS
The Medical Devices Directive 93/42/EEC states that product provider
must ensure that all personnel handling the product are provided with the
operating instructions & performance data.
Do not use the product without properly familiarization of the product
and its safe operation as defi ned in this Instruction for use. Ensure user
is aware of particular information and knowledge required for the gas
in use.
-20/+60°C
10/100%
600/1200 mbar
4/63
STORAGE AND TRANSPORT
CONDITIONS
-30/+60 °C
10/100%
60/1200 mbar
+N
O is
2
2