6.1.4. Functional checks before use
Ensure the fl ow control knob is on the "ZERO" position.
Ensure the cylinder valve is open – in the "ON" position.
Check that the gauge indicates pressure/contents. If the pointer reaches
the red area send the cylinder for the fi lling
For regulators fi tted with a fl ow-metering device check that there is gas
EN
fl ow at each setting (for instance, by listening for the sound of gas fl ow
or checking presence of bubbles in a humidifi er).
Turn off the shut off cylinder valve by turning the hand wheel in a clock-
wise direction to the stop position. Do not use excessive force.
Reset the fl ow control knob to on the "ZERO" position once the gas fl ow
stops and the regulator is vented..
For regulators fi tted with a pressure outlet, ensure it is functional by con-
necting and disconnecting a male QC probe.
6.2. USER OUTLET S CONNECTION & USE
6.2.1. List of recognised accessories
To be connected to the fl ow outlet:
Humidifi ers, breathing masks or cannulas, gas savers, nebulizers.
To be connected to the pressure outlet:
Flexible hoses, fl ow meters, Venturi suction ejectors.
At regulators fi tted with pressure outlet together with ejector outlet don´t
use quick coupler and ejector outlet in the same time. Especially when
inlet pressure is bellow 50 bar it may negatively a ect performance of
the regulator.
Before connecting any accessory or medical device to the regulator, al-
ways check that it is fully compatible with the product connection fea-
tures & the product performances.
6.2.2. Pressure outlet connection
PRESSURE OUTLET I
Ensure the male quick coupler is compatible with the pressure outlet
feature.
Connect the male quick coupler.
Check if the male quick coupler is fully engaged.
Regulator with threaded connector as pressure outlet shall be only an
integral part of medical equipment. Do not use it for other purposes!
PRESSURE OUTLET II
Ensure the counterpart is compatible with the pressure outlet features.
Screw the counterpart.
Check the counterpart is fully screwed.
8/63