Apéndice A
European Directive Information
INFORMACIONES SOBRE
CERTIFICACIONES
Este apéndice reúne informaciones relativas a los certificados que cubren el equipo. Su contenido
fue generado por el Equipo de Certificación de Smar Equipamentos Industriais Ltda.
European Directive Information
Authorized representative in European Community.
Smar Gmbh-Rheingaustrasse 9-55545 Bad Kreuzanach
This product complies with following European Directive:
EMC Directive (2004/108/EC) –"Electromagnetic Compatibility"
The EMC test was performed according to IEC standard: IEC61326-1:2005 and IEC61326-2-3:2006
Tests performed using twisted pair wire.
If using shielded cable, keep the shield insulated at the instrument side, connecting the other one to
the ground.
Immunity test requirements for equipment intended for use in industrial locations.
(Table 2, clause 6, IEC61326-1:2005)
PORT
Phenomenon
Electrostastic discharge
(ESD)
Enclosure
EM Field
Rated power frequency
magnetic field
Burst
I/o signal
Surge
/control
Conducted RF
Emission requirements (clause 7, IEC61326-1:2005).
According to the standards: CISPR11, IEC61000-3-2, IEC61000-3-3
PORT
TEST
Conducted
DC supply
Emission
Radiated
Housing
Emission
ATEX Directive (94/9/EC) – "Electrical equipment and protective system intended for use in
potential explosive atmospheres"
According to the standards: IEC60079-0:2006, IEC60079-1:2007, IEC60079-7:2001, IEC60079-
11:2007, IEC60079-15:2005
Notified body Nemko for QAM and QAR number CE0470
(Certification in process, declaration not issued yet )
Basic
Standard
IEC61000-4-2
IEC61000-4-3
IEC61000-4-8
IEC61000-4-4
IEC61000-4-5
IEC61000-4-6
Standard
Frequency Range
IEC 61000-3-2
0 to 2KHz
IEC 61000-3-3
150KHz to 500KHz
CISPR11
500KHz to 5MHz
Class A
5MHz to 30 MHz
30MHz to 230 MHz
CISPR11
Class A
230MHz to 1000MHz
Test Value
6KV/8KV contact/air
10V/m (80 to 1GHz)
3V /m (1,4GHz to 2GHz)
30A/m
1KV(5/50nS, 5KHz)
1KV (line to ground)
3V (150KHz to 80MHz)
Test Value
(not apply for transmitters)
79dB(μv) QP, 66dB(μV) AV
73dB(μV) QP, 60dB(μV) AV
73B(μV) QP, 60dB(μV) AV
40dB(μV) QP (A=10m)
47dB(μV) QP (A=10m)
Performance
Criteria
B
A
A
B
B
A
A.1