Preparation for measurement

The equipment trolley allows the bed scale to be used on a mobile basis, but the scale
can also be permanently installed.
– Move the equipment trolley next to the
bed.
– Press the brakes down onto the cas-
tors to lock the trolley in position.
– Remove one measuring foot and place
it next to one of the bed castors. Fold
out the actuating lever.
To unlock the lever, press down the
round knob.
– Keep the actuating lever at an angle of
approx. 45° to the floor and push the
measuring foot up against the bed.
The oscillating arm should now be next
to the bed castor.
– Now push the measuring foot further
forward – holding the oscillating arm
firmly in the other hand.
The oscillating arm is now under the
bed castor.
– Press the actuating lever downwards.
The measuring foot moves under the
castor, lifting the bed.
The floor under the measuring foot
should be level. You must make sure
that the connecting cable is not
caught under the measuring foot, oth-
erwise faulty measurements may re-
sult.