EMC DECLARATION FOR MYSCALE ANALYZER BW-SC2 AND MYSCALE ANALYZER XL BW-SC6
GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC IMMUNITY
THE BW-SC2 AND BW-SC6 ARE INTENDED FOR USE IN THE ELECTROMAGNETIC ENVIRONMENT SPECIFIED
BELOW. THE CUSTOMER OR THE USER OF THE BW-SC2 AND THE BW-SC6 SHOULD ENSURE THAT THEY
ARE USED IN SUCH AN ENVIRONMENT.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
48
BewellConnect - BW-SC2 / BW-SC3 / BW-SC4 / BW-SC6 - User Manual - 082017v1
IEC 60601 test
level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
±2 kV for pow-
±2 kV for pow-
er supply lines
er supply lines
±1 kV for input/
±1 kV for input/
output lines
output lines
±1 kV differen-
±1 kV differen-
tial mode
tial mode
±2 kV common
±2 kV common
mode
mode
<5 % Ur (>95
<5 % Ur (>95
% dip in Ur) for
% dip in Ur) for
0,5 cycles
0,5 cycles
40 % Ur (60 %
40 % Ur (60 %
dip in Ur) for 5
dip in Ur) for 5
cycles
cycles
70 % Ur (30 %
70 % Ur (30 %
dip in Ur) for 25
dip in Ur) for 25
cycles
cycles
<5 % Ur (>95
<5 % Ur (>95
% dip in Ur) for
% dip in Ur) for
5 sec
5 sec
3 A/m
3 A/m
Compliance
Electromagnetic environment –
level
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative hu-
midity should be at least 30 %.
Main power quality should be that
of a typical commercial or hospital
environment.
Main power quality should be that
of a typical commercial or hospital
environment.
Main power quality should be that of
a typical commercial or hospital en-
vironment. If the user of the BW-SC2
and the BW-SC6 requires continued
operation during power mains inter-
ruptions, it is recommended that the
BW-SC2 and the BW-SC6 be powered
from an uninterruptible power sup-
ply or a battery.
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commer-
cial or hospital environment.
guidance