Display
Automatic Calibration (Calimatic)
Action
Calibration with the first buffer is
terminated. Remove the sensor
and temp probe from the first
buffer solution and rinse them
thoroughly.
• One-point calibration:
End with cal.
Slope [%] and asymmetry
potential [mV] of the sensor
are displayed.
Proceed with enter.
• Two-point calibration:
Immerse sensor and tempera-
ture probe in the second buffer
solution.
Start with enter.
Retract sensor and temp probe
out of second buffer, rinse off,
re-install.
Repeat calibration: cal
End calibration: enter
pH value and "Hold" are dis-
played alternately. Sensoface
active, "enter" blinks.
Proceed with enter.
Hold is deactivated after 20 s.
Remark
For one-point
calibration only:
The calibration
process runs again as
for the first buffer.
The slope and asym-
metry potential of
the sensor (based on
25 °C) are displayed.
Confirmation
prompt.
11