SM101/SM150/SM175&TM121/TM125
DISPLAY LOGGERS OPERATION
Quick Start
• Install the DicksonWare Software. Press the "Features" button for features and detailed operation instructions (Help).
• Connect the cable (supplied with the software) to the logger and to a working serial port on your computer. If
the port is not COM1 refer to 'Setting up your Computer's Serial Port" in the "Getting Started" chapter in the
DicksonWare Help.
• Press the "Setup" button and then press the "Clear" button.
• The logger is now sampling and ready for use. The default sample interval is 2 minutes (5 minutes for
SM150/SM175).
• For SM150/SM175: connect Thermocouple Probes.
Specifications
Temperature Range (Logger): -30 to +122°F, (-34 to
+50°C)
Temperature Range (Probe):
TM125: -30 to +185ºF (-34 to +85ºC)
SM150/SM175: -300 to +2000ºF (-184 to +1093ºC);
-300 to +400ºF (-184 to +204ºC) with supplied probe
Temperature Sensor: Thermistor; K-type Thermocouple
(SM150/SM175)
Temperature Accuracy: ±1.8°F, (±1°C)
SM150 /SM175(Logger Only) Temperature Accuracy:
±1.8°F, (±1°C)
Humidity Range: (TM models) 0 to 95% RH
(non-condensing)
Humidity Sensor: (TM models) Capacitive
Humidity Accuracy: (TM models) ±2% from 0 to 90%
Operating Instructions
Note: Dew point available on TM models only. Also, SM150/SM175 displays "Prob" when the probe is unplugged.
{MIN/MAX}
Pressing this key will start the Min/Max display cycle.
{F/C}
Pressing this key toggles between Fahrenheit and Celsius on the unit's display. Holding this key down for 2 seconds
adds or removes calculated Dew Point temperature readings in the display rotation. If the 'Dew Point" is less than
32°F (0°C), it is a frost point and FP will be displayed in the lower right instead of dP.
{Display On}
Pressing this key turns the display on for approximately 6 seconds when the unit is running on its battery. When the
unit is running on AC power the display is always on.
Display on Continuous
Hold the {F/C} and {Display On} keys down together until the display comes on and cont. shows in the lower
left, then release Display On first. To get out of continuous mode press Display On. Note: Continuous on mode will
decrease battery life. When the optional plug in power supply is used the display is always on.
Clear Min/Max values
Holding the {MIN/MAX + F/C} keys down together until clr- appears on display clears the stored minimum and
maximum values. The Min and Max displayed by the logger will be the minimum and maximum values sensed since
it was last cleared. Dicksonware will show the minimum and maximum of the data set downloaded. These could
be different. For Example: Your logger is set for a 2 minute sample which is good for about 5 1 / 2 days and is in
wrap mode. After 3 days you clear Min/Max. After another 3 days you download the logger. The Min/Max displayed
by the logger will be the Min/Max for the last 3 days. The Min/Max shown by Dicksonware99 will be the Min/Max
for the last 5 1 / 2 days. Dew point Min/Max in the logger is not updated unless dew point is enabled (See the F/C
key instructions).
Battery Replacement
To replace the battery (9V alkaline) remove the 4 screws on the back of the case.
Factory Service and Returns
Contact the factory for a Return Authorization (RA) Number before returning any instrument. The model number,
serial number and a purchase order number will be requested before an RA number is issued.
Notice: Dickson shall not be liable for consequential or incidental damages resulting from
failure or malfunction of its products.
Dickson manufactures a complete line of temperature, humidity, air quality
& pressure instruments. Send or call for a free catalog of Dickson products.
Phone: 630.543.3747• Fax: 630.543.0498 • E-mail: DicksonCSR@Dicksonweb.com
DICKSON
www.dicksonweb.com
930 South Westwood Avenue • Addison, IL 60101-4997
Revised 3/01
at 73°F
Weight: Approximately 7 oz.
Display: 4 digit LCD
Resolution: 0.1°F/°C, 0.1%RH (TM models)
Display Update Rate: Approximately every two
seconds
Power Supply: One 9V alkaline battery (incl.)
Battery Life: Approximately 1 year at default sample
interval
Dimensions: 1.45" x 5.41" x 4.5"
(36.8mm x 137.4mm x 114.3mm)
Memory: 16,128 samples with 1 channel enabled, 8064
sample sets with 2 channels enabled

Publicidad

loading

Resumen de contenidos para Dickson SM101

  • Página 1 Contact the factory for a Return Authorization (RA) Number before returning any instrument. The model number, serial number and a purchase order number will be requested before an RA number is issued. Notice: Dickson shall not be liable for consequential or incidental damages resulting from failure or malfunction of its products.
  • Página 2 Se requerirá el número de serie, número de modelo y número de orden de compra antes de emitir cualquier número RA. Notificación: Dickson no será responsable por daños incidentales o indirectos resultantes de la falla o mal funcionamiento de sus productos.
  • Página 3 Avis: Dickson ne sera responsable d'aucuns dommages indirects ou accessoires résultant d'une défaillance ou d'un défaut de fonctionnement de ses produits.
  • Página 4 DICKSON CALIBRATION SERVICES Ask about our Calibration Reminder Service – IT’S FREE! Call 800.323.2448 DICKSON www.dicksonweb.com Faites vos achats en ligne à www.dicksonweb.com Téléphone: 630.543.3747 Télécopieur: 630.543.0498 Courriel: DicksonCSR@Dicksonweb.com 930 South Westwood Avenue Addison, IL 60101-4997 Révisé 3/01...

Este manual también es adecuado para:

Sm150Sm175Tm121Tm125