Mettler Toledo 8450 Manual De Servicio página 63

Báscula e impresora
Ocultar thumbs Ver también para 8450:
Tabla de contenido

Publicidad

Mercury-PC Setup Main Menu
Use a communications program such as Windows® HyperTerminal or ProComm® to
communicate with the Mercury-PC. The communications protocol for the Mercury-PC is:
9600, N, 8, 1.
With the cable plugged in, the scale power on, and your communications software
running and properly configured (9600, N, 8, 1), use a tool such as a paper clip to
press the setup switch in the access hole (see below). The set up Main Menu should
display. If it doesn't, press the up arrow key a few times.
Mercury-PC serial number 4267
Version 4.1 Build MTS24 EA2 Ethernet HW address 00:40:96:14:70:aa
Use arrow keys, or Ctrl-N and Ctrl-P to move selector bar.
Press Enter to make selection.
Initialize the Mercury-PC
For new scale setup, or anytime a radio is replaced, first select (1) "Reset configuration
to default", then select (2) "Reset the Mercury-PC". This will initialize the adapter and
configure it for the correct radio. Then (3) press the setup switch again to re-enter the
setup menu.
Mercury-PC serial number 4267
Version 4.1 Build MTS24 EA2 Ethernet HW address 00:40:96:14:70:aa
1
2
Use arrow keys, or Ctrl-N and Ctrl-P to move selector bar.
Press Enter to make selection.
3
Mercury-PC
Setup Switch
Capítulo 3: Configuración de la unidad
Nomadic Communications, Inc.
MAIN MENU
---------
Resume operation
Edit configuration
View configuration for capture
Reset configuration to default
View forwarding database
View system event log
Clear system event log
View RSSI Information
Reset the Mercury-PC
Mercury-PC Adapter Setup Screen
Nomadic Communications, Inc.
MAIN MENU
---------
Resume operation
Edit configuration
View configuration for capture
Reset configuration to default
View forwarding database
View system event log
Clear system event log
View RSSI Information
Reset the Mercury-PC
Ethernet RF
(6/00)
3-27

Publicidad

Tabla de contenido
loading

Tabla de contenido