Relative button
The relative measurement feature allows you to make measurements relative to a stored reference
value. A reference voltage, current, etc. can be stored and measurements made in comparison to that
value. The displayed value is the difference between the reference value and the measured value.
1
Perform any measurement as described in the operating instructions.
2
Press the RELATIVE button to store the reading in the display and the "REL" indicator will
appear on the display
3
The display will now indicate the difference between the stored value and the measured value.
4
Press the RELATIVE button to return to normal operation.
AC or DC Voltage Measurements
1
Insert the black test lead banana plug into the negative COM jack and the red test lead banana
plug into the positive V jack.
2
Turn the rotary switch to the VDC/AC position.
3
Press the MODE button to select AC or DC Voltage
4
Touch the test probes to the circuit under test and read the voltage on the display.
AC or DC Current Measurements
Caution: Do not make current measurements on the 20A scale for longer than 30 seconds
every 15 minutes. Exceeding 30 seconds may cause damage to the meter and/or the test leads.
1
Insert the black test lead banana plug into the negative COM jack and the red test lead banana
plug into the: