Operation
With sufficient pressure, operation of the OneFlow
is completely automatic. Dependable operation involves only
monitoring of outlet pressure, periodic filter changes and service
documentation.
Maintenance
Routine maintenance of your OneFlow
filter cartridge changes and/or replacement of sump O-rings. If the
system sizing recommendations have been followed, the OneFlow
system will provide a six-month filter replacement interval on most
tap water for the OF120RC cartridge. The OneFlow
OF120RM should last one year.
3.
2.
4.
System
®
system involves periodic
®
®
Cartridge
®
6.
7.
8.
Filter Change Frequency
The filters should be changed in response to the following conditions.
OF120RC
• 6 months since installation or last filter change
• Reduced water flow
OF120RM
• 12 months since installation or last filter change
If filter change frequency is less than six months due to pressure
drop, it may be necessary to add additional prefiltration or evalu-
ate system sizing recommendations.
Filter Cartridge Replacement Procedure
IMPORTANT
Determine whether all equipment connected to the OneFlow
system must be turned off prior to shutting off water supply from
filters.
1. If required, turn off equipment.
2. Turn OFF water to OneFlow
3. Press the red button to release pressure.
4. Remove housing(s) - use filter wrench if necessary.
5. Clean inside of housing sumps with warm water. If desired,
disinfect housings using a teaspoon of household bleach.
Add to filter bowl and add water. Let stand 5 minutes and
then discard.
6. Insert new cartridges into filter housings. Match cartridge
model numbers to model numbers on bracket.
7. Make certain the O-ring is properly positioned and reinstall
filter housings (hand tighten only). Check O-ring for damage
and replace if damaged or distorted.
8. Slightly open the inlet ball valve; push the red pressure
relief button to release trapped air until a small amount of
water comes out - release the red button and fully open
the ball valve.
9. Open the flush valve downline from the filter housing (if
equipped) and flush the new cartridges to drain or bucket
for two minutes or until water runs clear. If no flush valve is
present, disconnect line from equipment to flush to drain.
10. VERY IMPORTANT: With water supply inlet valve OPEN and
water flow confirmed, turn on connected equipment. Failure
to supply water to equipment may cause serious damage.
11. Record filter change on the service log.
5
system by closing Inlet Ball Valve.
®
®