Orbeco SC450 Manual De Instrucciones página 9

Ocultar thumbs Ver también para SC450:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Methods
Notes:
1.
Vial cleaning:
As many household cleaners (e.g. dishwasher detergent) contain reducing substances,
the subsequent determination of Chlorine may show lower results. To avoid any
measurement errors, only use glassware free of Chlorine demand.
Preparation: Put all applicable glassware into Sodium hypochlorite solution (0.1 g/l) for
one hour, then rinse all glassware thoroughly with deionized water.
2.
For individual testing of free and total Chlorine, the use of different sets of glassware is
recommended (EN ISO 7393-2, 5.3).
3.
Do not use the same sample vial for free and total chlorine without thoroughly rinsing
the vial between the two different tests.
4.
Preparing the sample:
When preparing the sample, the lost of Chlorine, e.g. by pipetting or shaking, must be
avoided. The analysis must take place immediately after taking the sample.
5.
The DPD color development is carried out at a pH value of 6.2 to 6.5. The reagents
therefore contain a buffer for the pH adjustment.
Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7 before
the reagent is added (use 0.5 mol/l Sulfuric acid resp. 1 mol/l Sodium hydroxide).
6.
Exceeding the measuring range:
Concentrations above 2 mg/l Chlorine can lead to results showing 0 mg/l. In this case,
the water sample must be diluted with water free of Chlorine and the measurement
repeated.
7.
Oxidizing agents such as Bromine, Ozone etc. interfere as they react in the same way as
Chlorine.
Reagent re-order information
DPD Free Chlorine Reagent, 10 ml size, 100 pack
DPD Free Chlorine Reagent, 10 ml size, 1,000 pack
DPD Total Chlorine Reagent, 10 ml size, 100 pack
DPD Total Chlorine Reagent, 10 ml size, 1,000 pack
For additional accessories, see page 17.
SC450_V1_11.2010
RP010-0
RP010-3
RP012-0
RP012-3
9

Publicidad

Tabla de contenido
loading

Productos relacionados para Orbeco SC450

Este manual también es adecuado para:

Hellige sc450

Tabla de contenido