en Instructions for Use
SPIRO HUMIDIFIER
Single patient use only
The HUMIDIFIER has an exchangeable filter and is used on a
tracheostomy cannula to take advantage of the heat and moisture that
is exhaled. It should be replaced with a new unit after 29 days of use, or
more often if necessary.
N.B. THE HUMIDIFIER MUST NEVER BE USED
WITHOUT THE FILTER
DAILY CHECKS: Check humidifier operation daily and inspect for
damage. Replace humidifier with new unit in case of damage or defective
operation.
CLEAN the valve HOUSING (3) daily. Do not use sharp tools. Use
the fingers or PLASTIC TWEEZERS (P/N 705, sold separately) to
remove/re-insert the filter. The valve housing and COVER (5) shall be
air dried after washing with luke-warm water and mild fragrance-free
detergent. Other detergents or solvents must not be used. Always check
the operation again after cleaning and check that the air passages are free.
Do not wash and re-use the filter!
REPLACE THE FILTER (4) (P/N 740, sold separately) as required,
at least twice every 24 hours. Replacement may be required more often,
perhaps several times per hour, if a build-up of mucus or phlegm is
present. Ensure that the filter is seated evenly against the bottom of the
filter housing, and that the air flow through the filter is not obstructed by
phlegm or other substances.
A HEAT SHIELD/SHOWER PROTECTOR (6) (P/N 703, sold
separately) should be used as required with the inhalation opening facing
downward.
An ADAPTER (2) (P/N 707, 708, sold separately) is only required for
connection to a cannula (1) without an ISO 15 connector (taper).
SUPPLEMENTAL OXYGEN can connect via the Ø5.5mm nipple on
the valve housing.
702