Symptom
A. More Volume/
Output than
Displayed or
Registered
B. Less Volume/
Output than
Displayed or
Registered
SPECIFICATIONS
Accuracy*: ±0.5% of the true flow at any
point within the linear range of the
turbine flowmeter.
Linearity*: ±0.5% of the mean K-factor
throughout the linear range.
Repeatability*: ±0.1% of the indicated
flow throughout the linear range.
Response Time:
½ in. to 2 in: 2-5 milliseconds for step
change in flowrate.
3 in: 10-25 milliseconds for step
change in flowrate.
Frequency Output: (over the linear range)
½ in. (051):
½ in. to 2 in:
3 in:
8
TROUBLESHOOTING
Probable Cause
1. Rotor may drag due to
foreign matter obstruction.
2. M a g n e t i c p i c k u p n o t
screwed down all the way
into the turbine flowmeter
body. This causes it not to
detect all the rotor blades
as they pass.
3. Turbine flowmeter installed
backwards.
4. Turbine flowmeter rotor
installed backwards.
5. K-factor is too high in elec-
tronic/readout device.
1. Caused by trapped air in
the process line.
2. K-factor is too low in elec-
tronic/readout device.
125 - 1000 Hz
100 - 1000 Hz
50 - 500 Hz
Solution
Check for debris inside the
meter. Clean and reassemble.
Screw the magnetic pickup all
the way down into the turbine
flowmeter body. Hand-tighten
only.
Install the turbine flowmeter in
accordance with the process
flow direction.
Install the turbine flowmeter
rotor in accordance with the
process flow direction.
Verify K-factor used. K-factor
should be decreased.
Install an air eliminator up-
stream of turbine flowmeter.
Verify K-factor used. K-factor
should be increased.
Voltage Output:
½ in. to 2 in:
Approx. 100 mV @ 100 Hz to
1.0 Volt (r.m.s.) @ 1000 Hz
3 in:
Approx. 80 mV @ 50 Hz to
800 mV (r.m.s.) @ 500 Hz
Maximum Overange: to 125% of maxi-
mum flowrate for intermittent periods.
Maximum Pressure:
½ in. to 2 in:
3 in:
Operating Temperature:
GIP, GBP, GIT & GBT:
-74°C - +107°C
GIT & GBT High Temperature:
-268°C - +427°C
* Data based on water calibration @ 77°F (25°C).
170 bar (2500 PSI)
136 bar (2000 PSI)