Before conducting any cleaning operation, SWITCH the unit OFF and disconnect the power supply.
In order to ensure optimal hygiene, you are advised to clean the machine for each new patient.
Use a DISINFECTANT (PROPANOL/ISOPROPANOL or ALDEHYDE-based solution). Spray the disinfectant on the SURFACES (plastic shells and metal components).
In order to ensure optimal hygiene, you are advised to clean the covers for each new patient. All the consumables enable hazard-free disposal.
a • Packing: Packing must be separated from the components plastic and paper/cardboard and given to the specific sites from recycling.
b • KINETEC PATIENT PAD KIT: To clean with a product of disinfection then to give it to the specific sites of recycling.
c • Unit: It contains electronic components, cables, aluminium, steel and plastic parts. When the splint is not operational any more, to dismount and separate in groups
from materials and to give them to correct unit of recycling or to turn over the machine to Kinetec for destruction.
Product:
Weight:
12Kg (24 pounds)
Splint dimensions:
95cm (37 inches) x 33cm (13 inches) x 33cm (13 inches)
Angular limits:
-10° to 120°
Speed:
from 45 to 155° per minute
Patient height:
full leg: 71 to 99 cm (28 to 39 inches)
Tibia: 38 to 53 cm (15 to 23 inches)
Femur: 33 to 46 cm (13 to 18 inches)
Electricity:
Power supply: 100-240 V~
Frequency: 50-60 Hz
Power consumption: 50 VA
Device of type B class I
IP 20.
Fuse T 750mA 250V 6.3x32mm (KINETEC order: 4610007434)
Environment
- Storage/transport conditions:
Temperature: -40 to 70°C / -40 to 160°F
Relative humidity: up to 90%
- Operating conditions:
Room temperature: 10 to 40°C / 50 to 105°F
Relative humidity: up to 80%
Product information
Cleaning
Elimination and recycling
Technical specifications
GB