6.4.
Maintaining a constant water level / Counter cooling
Caution:
Securely attach all tubing to prevent slipping.
Observe the laws and regulations of the water distribution company valid in
the location where the unit is operated.
For cooling tasks near the ambient air temperature the liquid
level/cooling set can be used for counter cooling.
By special pipe routing, cool faucet water is continuously supplied to
the water bath, while at the same time, the heated water is drained
via the overflow connection of the Level/Cooling set.
A specific water flow rate of 100 ml/minute is sufficient to
compensate for the characteristic temperature.
Use of the liquid level/cooling set for a continuous supply of
faucet water:
1. to keep the water level constant, especially for
applications up to the boiling point (supply of faucet
water only in the amount of evaporation losses).
2. for counter cooling of cooling tasks near the ambient
surrounding temperature (cool faucet water is
continuously supplied to the water bath, while at the
same time, the heated water is drained via the overflow
connection of the liquid level/cooling set).
Liquid level/cooling set
Order number: 8 970 415
11 drainage screw on water bath
20 compensation reservoir
21 connecting sleeve
22 supply/drainage sleeve
23 overflow sleeve
24 adaptor screw for constant liquid level function
25 adaptor screw assy. for counter cooling function and
simultaneous constant liquid level control
26 adjuster screw for filling level adjustment
Water Baths