Application: The load cells may be used in multiple cell applications Class III and IIIL consistent with the model designations,
number of scale divisions, and parameters specified in this certificate. Load cells of a given accuracy class may be used in
applications with lower accuracy class requirements provided the number of scale divisions, the v
suitable for the application. The manufacturer may market the load cell with fewer divisions (n
those listed on the certificate. However, the load cells will come with the appropriate n
used.
Specific Capacities, n
and v
max
Model
Capacity
PR 6203 Series
* load cells tested
Identification: An adhesive identification badge located on the cell, states manufacturer name, model, serial number, accuracy class
and rated capacity. Other pertinent information will be specified on the Calibration Certificate accompanying the cell.
Test Conditions: A 500 kg, 1000 kg, 2000 kg, 3000 kg and a 20 000 kg capacity load cell were tested by the NMi Certain B.V. at The
Netherlands facility. Testing was conducted in accordance with the OIML DoMC Mutual Acceptance Arrangement, signed by the
NCWM as a utilizing participant for load cell testing. Testing was conducted using deadweights as the reference standard. The load
cells were tested over a temperature range of -10
zero was measured and a time dependence (creep) test was performed. The barometric pressure test to determine sensitivity of the
load cell design to changes in barometric pressure was conducted. The data were analyzed for multiple load cell applications. OIML
R60 selection criteria were used to determine cells tested.
Evaluated By: S.J. Koeman, M.M.J. Meijer (NMi)
Type Evaluation Criteria Used: NIST, Handbook 44: Specifications, Tolerances and Other Technical Requirements for Weighing
and Measuring Devices, 2017. NCWM, Publication 14: Weighing Devices, 2017.
Conclusion: The results of the evaluation and information provided by the manufacturer indicate the device complies with applicable
requirements.
Information Reviewed By: J. Truex (NCWM)
Minebea Intec GmbH
Load Cell / PR 6203 Series
Values:
min
v
min
(kg)
500 kg*
0.20
1000 kg*
0.20
2000 kg*
0.20
3000 kg *
0.20
5000 kg
0.33
10 000 kg
0.67
20 000 kg*
1.00
25 000 kg
1.25
30 000 kg
1.50
40 000 kg
2.00
50 000 kg
2.50
60 000 kg
3.00
75 000 kg
3.75
100 000 kg
5.00
o
C to 40
Class III
Multiple Cell
n
max
2000
2000
5000
10 000
10 000
10 000
10 000
10 000
10 000
10 000
10 000
10 000
10 000
10 000
o
C with tests run on each cell at each temperature. The temperature effect on
Certificate Number: 17-111
value, and temperature range are
min
) and with greater v
max
and v
for which the load cell may be
max
min
Class IIIL
Multiple Cell
v
min
(kg)
0.20
0.20
0.20
0.20
0.20
0.22
0.33
0.42
0.50
0.67
0.83
1.00
1.25
1.67
Page 2 of 3
values than
min
n
max
2000
2000
10 000
10 000
10 000
10 000
10 000
10 000
10 000
10 000
10 000
10 000
10 000
10 000