Figure 3: Pneumatic Connections - Typical Use Cases
2.1 Supply Equipment
Pneumatic supplies should have isolation and venting valves and, where necessary,
conditioning equipment. The positive pressure supply should be regulated to between 110%
of the full-scale pressure range and MWP stated on the control module.
To protect the control module (Refer to Figure 5, Figure 6, Figure 7, item 8), from over-
pressure a suitable protection device (such as a relief valve or bursting disc) must be fitted to
limit the applied supply pressure to below the MWP (Refer to K0443 User Manual).
On instruments without a negative supply, the positive pressure discharges from the system
to atmosphere through the negative supply port. Fit the diffuser to the negative port to
diffuse airflow.
During system pressure vent operations, the pressure discharges from the system to
atmosphere through the negative supply and vent ports. Fit a diffuser to the vent port to
diffuse airflow.