Hotplates & Stirrers
UC150, UC151, UC152, UC152D
US150, US151, US152, US152D
Introduction
Thank you for purchasing this Stuart product.
To get the best performance from the
equipment, and for your own safety, please
read these instructions carefully before use.
Before discarding the packaging check that
all parts are present and correct.
This equipment is designed to operate under the
following conditions:
❖ For indoor use only
❖ Use in a well ventilated area
❖ Ambient temperature range +5°C to +40°C
❖ Altitude to 2000m
❖ Relative humidity not exceeding 80%
❖ Mains supply fluctuations not exceeding 10%
of nominal
❖ Overvoltage category II IEC60364-4-443
❖ Pollution degree 2 IEC664
❖ Use with a minimum distance all round of
200mm from walls or other items
If the equipment is not used in the manner
described in this manual and with accessories
other than those recommended by Cole-Parmer,
the protection provided may be impaired.
Electrical Installation
THIS EQUIPMENT MUST BE EARTHED
Before connection please ensure that the line
supply corresponds to that shown on the
rating plate located on the base of the unit.
Power requirements:
Model
Wattage
UC150
500W
UC151
50W
UC152
550W
UC152D
550W
3
Model
Wattage
US150
700W
US151
50W
US152
750W
US152D
750W
There is a hot condition IEC socket at the rear of
the instrument for connection to the mains supply
(see figure 2). The unit is supplied with a mains
lead fitted with either a UK 3-pin or "Schuko"
2-pin hot condition IEC plug for connection to the
instrument.
Should the lead not be suitable for connecting to
the mains power supply, replace the plug with a
suitable alternative.
THIS OPERATION SHOULD ONLY BE
UNDERTAKEN BY A QUALIFIED ELECTRICIAN.
NOTE: Refer to the equipment rating plate to
ensure that the plug and fusing are suitable for
the voltage and wattage stated.
The wires in the mains cable are coloured as
follows:
BROWN - LIVE
BLUE - NEUTRAL
GREEN/YELLOW - EARTH
Should the mains lead need replacement, a cable
2
of 1mm
of harmonised code H05RR-F or
H05RN-F connected to an IEC hot condition plug
should be used.
IF IN DOUBT CONSULT A QUALIFIED
ELECTRICIAN
Safety Advice
❖ Do not use the instrument to heat inflammable
liquids.
❖ Never lift or carry the instrument until it has
been switched off and allowed to cool for at
least 30 minutes. The hot warning light will
give guidance.
❖ The unit should be carried using both hands
with the fingers under the side edges.
❖ Never move or carry the unit with containers on
the top plate or while still connected to the
mains supply.
❖ There is a danger of liquid spillage if containers
are over-filled and stirred at high speed. Always
build stirrer speed slowly and never stir more
rapidly than necessary.
❖ NEVER place a cold glass vessel onto a hotplate
which is already hot.
❖ NEVER use a retort rod longer than 600mm.