DROPOUT VOLTAGE
The load will cease to conduct if the applied voltage falls below the Dropout Voltage setting;
active in all modes except Constant Voltage. The Dropout Voltage setting is also the threshold for
the Slow Start facility and acts as an offset voltage in Constant Resistance mode.
Setting Accuracy:
Slow Start
If Slow Start is enabled, the load will not conduct any current until the source voltage reaches the
Dropout Voltage setting; it will then ramp the controlled variable up (in CC, CP and CG modes) or
down (in CR and CV modes) to the Level setting at a rate determined by the Slew Rate setting.
METER SPECIFICATIONS
Display Type:
Measured Values
Volts & Amps:
Watt & Ohms:
Voltage Accuracy:
Current Accuracy:
CURRENT MONITOR OUTPUT
Output Terminals:
Output Impedance:
Scaling:
Accuracy:
Common Mode Range:
REMOTE CONTROL
Digital Remote Interfaces
The LD400P model provides LAN, USB, GPIB and RS232 interfaces for full remote control.
LAN:
USB:
GPIB:
RS232:
External Control Input Characteristics
Connection:
Input Impedance:
Common Mode Range:
External Analogue Voltage Control
Operating Mode:
Scaling:
Accuracy:
Common mode rejection:
External Logic Level (TTL) Control
Operating Mode:
Threshold:
4
± 2% ± 20mV.
256 x112 pixel graphic LCD with white LED backlight.
Measured values of current through and voltage across the load.
Power and equivalent load resistance, calculated from Volts and Amps.
± 0·1% ± 2 digits.
± 0·2% ± 3 digits.
4mm safety sockets on front panel or terminal block on rear panel.
600Ω nominal, for >1MΩ load (e.g. oscilloscope).
50mV per Amp (4 Volts full scale).
± 0·5% ± 5mV.
± 3Vdc max. to load negative. A connection is required, see
Ethernet 100/10base-T connection with auto cross-over detection.
1.4 LXI Core 2011 compliant.
Standard USB 2.0 connection. Operates as virtual COM port.
Conforming to IEEE488.1 and IEEE488.2.
Capabilities: SH1, AH1, T6, L4, SR1, RL2, PP1, DC1, DT0, C0, E2.
Standard 9-pin D connection. Baud rate: 9600.
Terminal block on rear panel.
400kΩ each input to load negative.
± 100V to load negative.
The applied voltage sets the operating level within the selected range.
4 Volts full scale.
± 2% ± accuracy of selected range.
Better than –66dB.
The applied signal selects between Level A and Level B settings.
+ 1·5V nominal. A logic high selects Level B.