11.

Measurement

Important Guidelines

Reagents and dilution buffers can cause cauterization and other damage to health.
Samples (nucleic acids, proteins, bacteria cultures) can be infectious and cause serious
damage to health.
During sample preparation, measuring procedures and maintenance and cleaning
work, observe all local laboratory safety precautions (e.g. wear protective clothing and
gloves, use of disinfectant) regarding the handling of sample material.
Dispose of measuring solutions and cleaning and disinfectant materials in accordance
with the relevant local laboratory regulations.

Check the cuvettes

The cuvettes must be clear and there's no remains of the samples on the surface of it.
Only Quartz cuvettes are permitted to be used in the range of UV area.

Photometry

Photometry mode is used to measure the absorbance or transmissivity of the sample.
1.
Main interface, press the icon
Mode Switch measurement mode to %T, Abs or Energy.
Wavelength Set measurement wavelength.
Zero Do 0Abs/100%T.
Read Measure sample and record the result.
List View the result(s) list.
UP/Down Increase/Decrease the gain of signal. Only for Energy mode.
to start a Photometry application.