PT2026 Installation and Safety Manual
2-2 RATINGS
Power
Overvoltage
Fuse
Environment
Operating temperature
Storage / transport
temperature
Altitude
Relative humidity
Pollution
Magnetic environment
Electromagnetic
environment
v 1.0 r 4.0 – 02/2017
55 VA, 100 – 240 VAC, 50-60 Hz
See warnings below.
Accepts temporary overvoltages occurring on the
mains supply. Transient overvoltages up to the levels
of overvoltage category ll.
3.15 A (T), 5x20 mm, 250 V
See warning below.
Indoor use; no air inlet (IP 50)
10 – 40 °C
Deuterium probes: 0 – 80 °C
Other: -25 – 80 °C
≤ 2000 m
Maximum 80 % for temperatures up to 31 °C,
decreasing linearly to 50 % relative humidity at 40 °C
Pollution degree 2: only non-conductive pollution
occurs, except that occasionally a temporary
conductivity caused by condensation is to be
expected.
< 0.2 T (PT2026 Main Unit)
Equipment intended to be used in an industrial
electromagnetic environment, class A.
See warning below.