G. MAINTENANCE OF YOUR VIOS
NEBULIZER SYSTEM
FILTER
Your Vios
air filter is bright white when new. With use,
®
the filter will slowly turn grey, indicating the need for
replacement. The filter must be replaced immediately if it
gets wet or becomes clogged. A good rule to follow is to
replace your filter every 6 months – at the same time as
your PARI reusable nebulizer .
Removing: First make sure the Vios
turned off. Using a coin, remove the air filter cover by
turning it counter-clockwise (Fig.E). Pull the cap and filter
out, then remove the filter from the cap (Fig.F-G).
NOTE: THERE ARE TWO TYPES OF FILTERS THAT FIT
THE VIOS
COMPRESSOR. IF YOUR FILTER IS OPEN
®
ONLY ON ONE END AS SHOWN IN FIGURE F,
FOLLOW INSTRUCTIONS AS SHOWN IN FIGURE F
AND G. IF YOUR FILTER IS OPEN ON BOTH ENDS,
PRESS EITHER END OF THE FILTER INTO THE CAP.
Take a new filter and press into cap with the open end
visible. Place the cap and filter back into the front of
your Vios
. Use the coin to tighten by turning the cover
®
clockwise.
NOTE: DO NOT ATTEMPT TO WASH AND REUSE
AIR FILTERS. DO NOT USE SUBSTITUTE FILTER
MATERIAL COMPRESSOR DAMAGE MAY RESULT.
F
®
®
power switch is
Nebulizer Replacement Sticker
Suggested Placement
COMPRESSOR
Use a damp cloth to wipe the exterior of your Vios
housing and Wing Tip
CAUTION
Never spray fluid into the ventilation slots on your Vios
Fluid could cause damage to the electrical & compressor
parts and could lead to malfunction. In the event fluid
enters the unit, please contact PARI Patient Connection at
1.800.FAST.NEB (327.8632).
MAINTENANCE & REPAIR
Never operate your Vios
incompletely assembled or damaged. Use only original
PARI parts and attachments authorized by PARI for the
Vios
Aerosol Delivery System. Use of other parts and
®
attachments may damage the unit or cause improper
treatment. Do not open or disassemble the compressor
unit at any time.
NOTE: THERE ARE NO SERVICEABLE PARTS INSIDE
THE VIOS
AEROSOL DELIVERY SYSTEM. SERVICE BY
®
ANYONE VOIDS WARRANTY.
REMEMBER
REPLACE YOUR REUSABLE NEBULIZER CUP EVERY
SIX MONTHS
It is very important to replace your PARI reusable nebulizer every
6 months for the most effective treatments.
Advantages of Replacing Your Nebulizer Cup:
-Consistent medication delivery to the lungs
-Fast and efficient treatments
-Bacterial contamination can be reduced*
Included in this configuration is a reminder sticker. Place the sticker
that corresponds to six months from date of first use to the smooth
surface on the bottom of your nebulizer cup as indicated in this
picture. Sticker is dishwasher safe.
Allow two hours after label is applied before cleaning.
WARNING: The sticker may present a choking hazard to small
children-adult supervision is required.
* Lester, Mary K. Nebulizer Use and Maintenance by Cystic Fibrosis
Patients: A Survey Study. Respiratory Care 2004; 49(12): 1504-1508.
G