Output voltage
Test Frequency
Transformer rating
Output disconnection
Power supply
Fuses
Operation
Display
Measurement Accuracy
Controls
Safety
EMC
Operational uncertainties
Temperature Range
Humidity Range
Dimensions
Weight
Cleaning
Specification
0 - 60 kV rms
61,8 Hz
500 VA
Within 1 ms of detection of breakdown
115 V (99 - 132 V), 230 V (198 - 264 V) (switch selectable) 50/60 Hz, 80 VA
500 mA (T) HBC to IEC127/5 for 220/240 V supply
1 A (T) HRC to IEC127/5 for 110/120 V supply
Mains power cord fused plug (when applicable): 3 Amp to BS1362
Semi automatic with 0,5, 2 or 3 kV/s selectable rate of voltage rise
Pause function with automatic 1 minute timer for Withstand testing
0,0 - 60,0 kV LED digital display
HV on LED
Ramp rate LED
2% ± 3 digits
Start, Pause, Stop, Ramp Rate select and Power On/Off switch
The instrument meets the requirements of IEC 61010-1
In accordance with IEC 61326-1
Refer to www.megger.com
Operating:
0 to 40 ºC
Storage:
-40 to +70 ºC
Operating:
80% RH at 40 ºC
Storage:
93% RH at 40 ºC, 95% RH at 25 ºC
336 mm (13,2 in) (H) x 400 mm (5,7 in) (L) x 235 mm (9,3 in) (D)
17,5 kg (38,5 lb)
Wipe disconnected test set with a clean cloth dampened with soapy water or Isopropyl Alcohol (IPA)
11