Copy of the EC Declaration of conformity can be requested to Gima S.p.A. - Via Marconi,
1 - 20060 – Gessate, Milan (Italy)
NB: The measures and the curves are not valid in the event of medications supplied in high
viscosity suspension.
CLEANING OF THE APPLIANCE
The cleaning of the device shall be carried out with a soft and dry cloth and with non-abrasive
detergents and no solvents. Do not use excessively wet cloths as the contact of liquids with
the electrical parts of the appliance may cause malfunctions or may be hazardous.
While cleaning the device, make sure that no liquids get inside the appliance and
that the power outlet is disconnected.
Wait for the device to dry off before storing or using it again.
MAINTENANCE
The Corsia device does not have any part that requires maintenance and/or lubrication.
Nonetheless, it is necessary to carry out some checks to verify the function and security of
the appliance before each use. Before using the device, carry out disinfection as described in
the "CLEANING ACCESSORIES" section. Remove the appliance from the box and always
check that there is no visible damage; particular attention should be given to any cracks in
plastics which may leave some electrical components exposed. Check also the integrity of
the power cable which might have been damaged during the previous use.
Then connect the cable to the power network and switch it on. Close the air outlet with one
finger to make sure that the produced noise is regular and there is no malfunctioning.
Check that the nebulizer has not suffered any breakage during the previous use (it was
stored inappropriately or has suffered impacts). An easily replaceable protection fuse (F 1 x
1.6A L 250V) situated in the power plug on the back of the device protects the instrument.
For its replacement, always check the type and the specified range.
ENGLISH
MMAD = Mass Median
Aerodynamic
Diam-
eter
GSD
=
Geometric
Standard Deviation
18