R&S HZ-14
The near-field probe set comprises:
5
Magnetic probe 9 kHz to 30 MHz
5
Magnetic probe 30 MHz to 1 GHz
5
Electric probe 9 kHz to 1 GHz
5
Probe power adapter for electric probe
5
Preamplifier 9 kHz to 1 GHz for magnetic probes
5
Probe test jig
5
RF Cable
5
2 power cables for connection on R&S Test Receivers and Spectrum Analyzers
5
Carrying Case
Near-field Magnetic Probes
The near-field magnetic probes are passive. Any magnetic field (H) whose magnitude
varies with time and which passes through the end face of the probe tip generates a
voltage at the probe output.
Near-field Electric Probe
The near-field electric probe is an active probe with power supplied via the RF cable.
The probe power adapter includes an acoustic overload warning. The pointed tip of this
probe makes it especially useful for exactly locating sources of interference.
Probe Test jig
The probe test jig is used with a measuring receiver or spectrum analyzer and tracking
generator to verify that a magnetic probe has not been damaged in use.
Immunity Testing
Both magnetic probes being passive are bi-directional, i.e. power may be applied to the
probe connector to enable the sensing head to induce currents into conductors for
localized immunity testing.
Operating Manual 1026.7767.12 - 01
Introduction
4