Preliminary Operations; Sterilisation Temperature Programming; Sterilisation Time Programming - VWR VAPOUR-Line lite Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 50
body handle for proper cover and body alignment. Turn the cover in the direction indicated to close
until the cover handles are centred directly above body handles.
Do not rotate the cover beyond this point.
When the cover is rotated to the closed position, the air vent/cover lock passes under the locking
bracket. When the handles are aligned, one over the other, the pressure regulator is placed on the
vent pipe and heat is applied, pressure will then begin to build within the vessel.
At this point, the air vent/cover lock lifts and locks the unit. The cover will remain locked as long as
there is pressure in the vessel. When pressure is reduced completely, the air vent/cover lock drops
allowing the unit to be opened.

Preliminary operations

1) This steam steriliser must only be used for a maximum pressure of 1.08 bar (121 °C).
2) The steam steriliser should be positioned in an area where there is adequate means for
removal of steam.
3) Connect plug to mains supply.
4) Introduce distilled water or demineralised water into the sterilisation chamber to cover the
heating system. Using distilled water avoids the production of calcareous deposits on the
heating element. 20 minutes of sterilising will use approximately 500 ml of water.
5) Two displays are lit on time/temperature command panel.
6) Before programming, it is necessary to erase the protection of the parameters (safety keys),
see paragraph 16.
Sterilisation temperature setting
1) 5.1 Press 'ENTER' four times in quick succession to obtain 'SP1' on lower display.
2) 5.2 Select the sterilisation temperature using the arrows. The sterilisation temperature will be
shown on the upper display.
3) 5.3 Press 'ENTER' to save to memory.
Sterilisation time setting
1) Press 'ENTER'. 't in 1' appears on the lower display.
2) Select the sterilisation time using the arrows. The sterilisation time will be shown on the upper
display.
3) Press 'ENTER' to save to memory.
4) Starting values will be shown on the display after 30 seconds (vessel actual temperature and
the word 'OFF').
7