5 Maintenance and service
Generally, all moveable parts must be inspected visually and for functionality
every 6 months; all non-moveable structural parts every 12 months. If the
product is used for light sportive activities, the inspection intervals must be
reduced to a maximum of 3 months.
If necessary, clean the joint (not with compressed air!) and re-adjust.
In case of severe functional problems of the hydraulic components or
increased sound development, please return the knee joint to Streifeneder
ortho.production GmbH after prior notice.
Attention!
Extensive wear of the end-damper may lead to limited function.
Attention!
In order to eliminate sounds in the cosmetic cover, use silicone
spray 116P11; do not use talcum. Using talcum causes increased
danger of falling of the patient, as it withdraws grease from the
mechanic construction parts, which may cause malfunctions.
Using this product in connection with talcum will lead to loss of
all warranty rights against Streifeneder ortho.production.
Spray the silicone spray 116P11 directly onto the friction surfaces
of the cosmetic cover.
32/60